MediaClave overview
INTEGRA Biosciences
Page 48
6.4.2.3
Error messages during operation
Displayed Text
Description/Action
WARNING:
PLEASE CLOSE COVER!
Description:
Before starting operation, the
MediaClave verifies the sealing of the safety
cover. If the cover is opened, the user is
informed.
Reaction:
Close the cover.
WARNING!
PLEASE OPEN BOILER!
Description:
In mode Water bath, the Media-
Clave verifies the sealing of the vessel. If the ves-
sel is closed, the user is informed.
Reaction:
Open vessel.
SENSOR SHORT!
Description:
The MediaClave verifies before and
during operation the functionality of product and
coupling medium temperature sensors. The
‚SHORT-WARNING‘ is issued when the meas-
ured temperature is below -20°C.
Reaction:
Check sensor wiring and connector.
WARNING!
SENSOR SHORT!
PLEASE CHECK SENSORS!
ICE IN THE SYSTEM!
Description:
The MediaClave verifies before and
during operation the functionality of both temper-
ature sensors. The ‚ICE IN THE SYSTEM-
WARNING‘ is issued when the temperature is
below 2°C.
Reaction:
Check sensor and access
wire, use MediaClave only within the specified
environment conditions.
WARNING!
ICE IN THE SYSTEM!
PLEASE CHECK SENSORS!
INTERRUPTION TEMPERATURE
SENSOR!
Description:
The MediaClave verifies before and
during operation the functionality of product and
coupling medium temperature sensors. The
‚INTERRUPTION TEMPERATURE SENSOR-
WARNING‘ is issued, when temperature is over
300°C.
Reaction:
Check sensor and access
wire.
WARNING!
INTERRUPTION TEMPERATURE SENSOR!
PLEASE CHECK SENSORS!
TOO LITTLE HEATING WATER IN BOILER!
Description:
The MediaClave verifies before and
during operation the water level of the coupling
medium. If the level of the coupling media falls
below the minimum, the user is informed.
Reaction:
Terminate the program and refill. The
current program is blocked until the MediaClave
has cooled down.
WARNING!
WATER BELOW MINIMUM LEVEL!
WAIT UNTIL MACHINE HAS COOLED DOWN
PRESSURE TEST NOT SUCCESSFUL!
Description:
At start-up, a pressure test is
enforced.
Reaction:
Terminate program and
check lid seal and sterile filter.
ERROR!
SAFETY TEMPERATURE HEATER!
PLEASE CALL CUSTOMER SERVICE!
Description:
The MediaClave verifies before and
during operation the functionality of the overtem-
perature switch.
Reaction:
Call customer service
Table 6.4:
Operational Error Messages
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...