Resolving errors/Eliminating errors
FAM 5
en-US
Page 87 / 148
BEWA FAM5 4166129a en-us 2019-03-04.docx
2019-03-04
Resolving errors/Eliminating errors
DANGER
Electric shock
Danger of fatal injury
►
Any work involving electrical equipment may
only be done by a properly trained, certified
electrician.
If an error occurs, the following takes place:
•
The FAM switches to "ERROR" status
•
the FAM ceases operation
•
the fault indicator lamp flashes
•
The display flashes yellow.
Press the down
▼ arrow key to display the error message.
ERROR
until
until
------
OM
MAN.
SET
Dewater
Filter
Heat
ERROR
RESET with ENTER
23 Filter clogged
33 Vacuum column alarm
34 Drip tray full
NEXT ERRORS ▼▲
# 33
# 34
1
2
+518
3