ES User’s Guide
3-23
Making Measurements
Measuring S-Parameters using a Two-Port Calibration
5. Press
after connecting each standard.
6. Next, the instrument prompts you to connect three standards (open,
short, and load) to port 2. Instead of connecting the standards directly
to port 2, connect them to the open end of the cable. This end of the
cable will become the calibrated port 2 for measurements. See the
graphic above.
7. The analyzer will measure each standard and then calculate new
calibration coefficients. The message "
Calibration complete.
"
will appear for a few seconds when the analyzer is done calculating
the new error correction array.
NOTE
Changing sweep frequencies (and other source parameters) may affect
your calibration. See
Port 1: Open, Short, Load Connections
Port 2: Open, Short, Load Connections
Measure Standard
Содержание HP 8712ES
Страница 8: ...viii ES User s Guide ...
Страница 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Страница 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Страница 35: ...ES User s Guide 2 1 2 Getting Started ...
Страница 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Страница 53: ...ES User s Guide 3 1 3 Making Measurements ...
Страница 58: ...3 6 ES User s Guide Making Measurements Measuring Devices with Your Network Analyzer Figure 3 3 Block Diagram ...
Страница 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Страница 146: ...4 12 ES User s Guide Using Instrument Functions Using Markers Figure 4 6 6 dB Bandwidth Marker Search ...
Страница 151: ...ES User s Guide 4 17 Using Instrument Functions Using Markers Figure 4 11 Multi Notch Search Mode ...
Страница 153: ...ES User s Guide 4 19 Using Instrument Functions Using Markers Figure 4 12 Marker Statistics Function ...
Страница 155: ...ES User s Guide 4 21 Using Instrument Functions Using Markers Figure 4 13 Marker Flatness Function ...
Страница 157: ...ES User s Guide 4 23 Using Instrument Functions Using Markers Figure 4 14 RF Filter Statistics Function ...
Страница 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Страница 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Страница 183: ...ES User s Guide 4 49 Using Instrument Functions Customizing the Display Figure 4 22 The Display Annotation ...
Страница 187: ...ES User s Guide 4 53 Using Instrument Functions Customizing the Display Figure 4 23 Normal Display ...
Страница 188: ...4 54 ES User s Guide Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display ...
Страница 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Страница 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Страница 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Страница 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Страница 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Страница 283: ...ES User s Guide 7 3 Front Rear Panel Connectors Connectors Figure 7 1 Analyzer Connectors Front Panel ...
Страница 303: ...ES User s Guide 7 23 Front Rear Panel Line Module Figure 7 11 Power Cable and Line Mains Plug Part Numbers ...
Страница 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Страница 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Страница 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Страница 411: ...ES User s Guide 9 1 9 Specifications ...
Страница 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Страница 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Страница 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Страница 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...