ES User’s Guide
5-21
Optimizing Measurements
Measuring Devices with Long Electrical Delay
Measuring Devices with Long Electrical
Delay
When making narrowband-receiver measurements of devices with long
electrical delay, measured levels can be affected by the rate at which the
source is changing frequency. This sensitivity is related to the time
required for the source signal to travel through cables or devices which
are connected between the analyzer’s two test ports. Since the source
frequency is changing rapidly during a sweep, a long distance or delay
between the analyzer’s two test ports will mean that the signal arriving
at the input port will be of slightly lower frequency than the output
signal at the same moment in time. This effect is referred to as
"frequency shift." The amount of frequency shift is given by the following
equation:
The narrowband receiver at the analyzer’s input port is tuned to the
exact frequency being emitted at the analyzer’s output port, with an
input bandwidth determined by the system-bandwidth selection. If the
input signal is lower in frequency than the output signal, the
measurement of the input signal will be attenuated by the frequency
response of the system bandwidth. The amount of attenuation increases
as the amount of frequency shift increases. The amount of attenuation
also increases as system bandwidth decreases.
The analyzer has been designed to minimize the effect of frequency shift
when a short cable is connected between the analyzer’s two test ports.
When a long cable (or a device with long electrical delay) is connected,
however, it is possible for the measurement to be affected, especially at
the analyzer's fastest sweep rates. If frequency shift is suspected, use the
following techniques to reduce its effect:
• Increase sweep time.
• Decrease frequency span.
• Select a wider system bandwidth.
• Use shorter cables to connect the DUT to the analyzer.
• Use broadband detection to completely eliminate the effect of
frequency shift.
frequency s hift
transit time
frequency span
sweep time
-----------------------------------------
×
=
Содержание HP 8712ES
Страница 8: ...viii ES User s Guide ...
Страница 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Страница 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Страница 35: ...ES User s Guide 2 1 2 Getting Started ...
Страница 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Страница 53: ...ES User s Guide 3 1 3 Making Measurements ...
Страница 58: ...3 6 ES User s Guide Making Measurements Measuring Devices with Your Network Analyzer Figure 3 3 Block Diagram ...
Страница 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Страница 146: ...4 12 ES User s Guide Using Instrument Functions Using Markers Figure 4 6 6 dB Bandwidth Marker Search ...
Страница 151: ...ES User s Guide 4 17 Using Instrument Functions Using Markers Figure 4 11 Multi Notch Search Mode ...
Страница 153: ...ES User s Guide 4 19 Using Instrument Functions Using Markers Figure 4 12 Marker Statistics Function ...
Страница 155: ...ES User s Guide 4 21 Using Instrument Functions Using Markers Figure 4 13 Marker Flatness Function ...
Страница 157: ...ES User s Guide 4 23 Using Instrument Functions Using Markers Figure 4 14 RF Filter Statistics Function ...
Страница 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Страница 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Страница 183: ...ES User s Guide 4 49 Using Instrument Functions Customizing the Display Figure 4 22 The Display Annotation ...
Страница 187: ...ES User s Guide 4 53 Using Instrument Functions Customizing the Display Figure 4 23 Normal Display ...
Страница 188: ...4 54 ES User s Guide Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display ...
Страница 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Страница 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Страница 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Страница 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Страница 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Страница 283: ...ES User s Guide 7 3 Front Rear Panel Connectors Connectors Figure 7 1 Analyzer Connectors Front Panel ...
Страница 303: ...ES User s Guide 7 23 Front Rear Panel Line Module Figure 7 11 Power Cable and Line Mains Plug Part Numbers ...
Страница 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Страница 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Страница 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Страница 411: ...ES User s Guide 9 1 9 Specifications ...
Страница 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Страница 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Страница 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Страница 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...