ES User’s Guide
5-17
Optimizing Measurements
Reducing Mismatch Errors
Reducing Mismatch Errors when Measuring
Both Reflection and Transmission
When you want to measure reflection and transmission simultaneously,
or without changing the test setup, perform a two-port calibration.
Though not as accurate as a two-port calibration, both a one-port
calibration for reflection and an enhanced response calibration for
transmission can be performed using the exact frequency parameters
that you will be using for the measurement. With this calibration choice,
it is recommended that you put a 6 to 10 dB pad directly on the output of
the DUT to improve measurement accuracy. Using both a one-port
calibration and an enhanced response calibration offers the advantage of
faster measurement speeds than when using a two-port calibration.
NOTE
Using an attenuator on the output of the DUT will reduce the system
dynamic range.
Содержание HP 8712ES
Страница 8: ...viii ES User s Guide ...
Страница 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Страница 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Страница 35: ...ES User s Guide 2 1 2 Getting Started ...
Страница 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Страница 53: ...ES User s Guide 3 1 3 Making Measurements ...
Страница 58: ...3 6 ES User s Guide Making Measurements Measuring Devices with Your Network Analyzer Figure 3 3 Block Diagram ...
Страница 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Страница 146: ...4 12 ES User s Guide Using Instrument Functions Using Markers Figure 4 6 6 dB Bandwidth Marker Search ...
Страница 151: ...ES User s Guide 4 17 Using Instrument Functions Using Markers Figure 4 11 Multi Notch Search Mode ...
Страница 153: ...ES User s Guide 4 19 Using Instrument Functions Using Markers Figure 4 12 Marker Statistics Function ...
Страница 155: ...ES User s Guide 4 21 Using Instrument Functions Using Markers Figure 4 13 Marker Flatness Function ...
Страница 157: ...ES User s Guide 4 23 Using Instrument Functions Using Markers Figure 4 14 RF Filter Statistics Function ...
Страница 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Страница 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Страница 183: ...ES User s Guide 4 49 Using Instrument Functions Customizing the Display Figure 4 22 The Display Annotation ...
Страница 187: ...ES User s Guide 4 53 Using Instrument Functions Customizing the Display Figure 4 23 Normal Display ...
Страница 188: ...4 54 ES User s Guide Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display ...
Страница 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Страница 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Страница 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Страница 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Страница 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Страница 283: ...ES User s Guide 7 3 Front Rear Panel Connectors Connectors Figure 7 1 Analyzer Connectors Front Panel ...
Страница 303: ...ES User s Guide 7 23 Front Rear Panel Line Module Figure 7 11 Power Cable and Line Mains Plug Part Numbers ...
Страница 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Страница 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Страница 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Страница 411: ...ES User s Guide 9 1 9 Specifications ...
Страница 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Страница 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Страница 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Страница 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...