SAMPle
The
SAMPle
subsystem provides commands to set and query the interval between
channel measurements (pacing) for each of the four Scan Lists.
Subsystem Syntax
SAMPle
:TIMer <
scan_list
>,<
interval
>
:TIMer? <
scan_list
>
SAMPle:TIMer
SAMPle:TIMer <
scan_list
>,<
interval
>)
sets the time interval between channel
measurements for the specified Scan List, or ALL four Scan Lists. When
SAMPle:TIMer LISTL
is specified,
<
interval
>
sets the time between channels for all
scan lists defined in
LISTL
, overriding any timer settings for individual Scan Lists.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
scan_list
>
discrete (string)
LIST1 | LIST2 | LIST3 | LIST4 | LISTL | ALL
none
<
interval
>
numeric (float 32)
(string)
1.0E-5-5 to 32.768E-3 | MIN | MAX
seconds
Comments
•
The minimum
<
interval
>
is 10
µ
seconds. The resolution for
<
interval
>
is .5
µ
s.
<
interval
>
may be specified in milliseconds (ms), or microseconds (us).
Note
If the A/D filter is on (
SENSe:FILTer:LPASs:STATe ON
) then the minimum
interval is 145
µ
sec.
•
If the Sample Timer interval multiplied by the number of channels in the specified
Scan List is longer than the Trigger Timer interval, at run time Error +3012,
“ Trigger too fast” will be generated.
•
This command will cause Error +3000, “ illegal while initiated” if trigger is initiated.
•
Related Commands:
SAMPle:TIMer?
•
*RST Condition: Sample Timer for all Channel Lists set to 1.0E-5 seconds.
Usage
SAMP:TIM LIST3,1E-3
Pace measurements at 1 millisecond
intervals for channels in Scan List 3.
SAMPle
226 HP E1313/E1413 Command Reference