[SENSe:]STRain:UNSTrained
[SENSe:]STRain:UNSTrained <
unstrained_v
>,(@<
ch_list
>)
specifies the unstrained
voltage value to be used to convert strain bridge readings for the channels specified by
<
ch_list
>
. This command does not control the output voltage of any source.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
unstrained_v
>
numeric (float32)
-16 t16
volts
<
ch_list
>
channel list (string)
100 - 163
none
Comments
•
Use a voltage measurement of the unstrained bridge sense channel to determine
the correct value for
<
unstrained_v
>
.
•
<
ch_list
>
must specify the channel used to sense the bridge voltage, not the
channel position on a Bridge Completion SCP.
•
Related Commands:
[SENSe:]STRain:UNSTrained?
,
[SENSe:]FUNCtion:STRain:
…
commands
•
*RST Condition: Unstrained voltage is zero.
Usage
STRAIN:UNST .024,(@100)
Set unstrained voltage for channel 0.
[SENSe:]STRain:UNSTrained?
[SENSe:]STRain:UNSTrained? (@<
channel
>)
returns the unstrained voltage
value currently set for the sense channel specified by
<
channel
>
. This command
does not make a measurement.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
channel
>
channel list (string)
100 - 163
none
Comments
•
Returned Value: Numeric value of unstrained voltage. The C-SCPI type
returned is flt32.
•
<
channel
>
must specify a single channel only.
•
Related Commands:
[SENSe:]STRain:UNSTrained
Usage
STRAIN:UNST? (@107)
Query unstrained voltage for channel 7.
Enter statement here.
Returns the unstrained voltage set by
STR:UNST.
[SENSe:]
Chapter 5
HP E1313/E1413 Command Reference 251
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com