0A 0004 (channel number)
0C 0008 (EU code for J thermocouple)
note 5
08 0010 (
ASSIGN
cmd)
notes 2,4
Continue for all 64 channels
DSP checks SCP Gains
Reg Offset (hex) Value (hex)
08 0820 (
SCPGAINS
cmd)
note 2
Calibrate SCP Channels
Reg Offset (hex) Value (hex)
08 1000 (
CARDCAL
cmd)
note 2
Clear Scan List 1
Reg Offset (hex) Value (hex)
08 0100 (
NEW
n cmd for Scan List 1)
note 2
Define Scan List 1
Perform for each channel you want in scan list 1
Reg Offset (hex) Value (hex)
0A 0000 (channel number)
0C FFFF (range code , FFFF=autorange)
0E 7 (data_flags, 7=EU converted to FIFO
and CVT)
08 0110 (
APPEND
cmd)
notes 2,4
Select Scan List 1
Reg Offset (hex) Value (hex)
10 0000 (See Scan Control Bits on page 343)
Arm the Module
Reg Offset (hex) Value (hex)
0A 0000
08 0400 (
ARM
cmd)
notes 2,4,6
Send Software Trigger
Reg Offset (hex) Value (hex)
26 0001