Chapter 1
Getting Started
About This Chapter
Except where noted, all references to the HP E1413 apply to the
HP E1313. This chapter will explain hardware configuration before
installation in a VXIbus mainframe. By attending to each of these
configuration items, your HP Scanning A/D Converter module will not need
to be removed from its mainframe later. Chapter contents include:
•
Configuring the HP Scanning A/D Converter
Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 15
•
Instrument Drivers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 29
•
About Example Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 29
•
Verifying a Successful Configuration . . . . . . . . . . . . . . . . . . Page 32
Configuring the HP Scanning A/D Converter Module
There are several aspects to configuring the module before installing it in a
VXIbus mainframe. They are:
•
Setting the Logical Address Switch. . . . . . . . . . . . . . . . . . . . Page 16
•
Installing HP E1313 Signal Conditioning Plug-ons . . . . . . . . Page 18
•
Installing HP E1413 Signal Conditioning Plug-ons . . . . . . . . Page 22
•
Disabling the Input Protect Feature . . . . . . . . . . . . . . . . . . . . Page 26
•
Disabling Flash Memory Access . . . . . . . . . . . . . . . . . . . . . . Page 26
For most applications you will only need to change the logical address
switch prior to installation. The other settings can be used as delivered.
Switch/Jumper
Setting
Logical Address Switch
24
Input Protect Jumper
Protected
Flash Memory Protect Jumper
PROG
Note
Setting the VXIbus Interrupt Level: The HP Scanning A/D Converters use
a default VXIbus interrupt level of 1. The default setting is made at power-on
and after a
*RST
command. You can change the interrupt level by executing
the
DIAGnostic:INTerrupt[:LINe]
command in your application program.