Comments
•
The
<
range
>
parameter must be within
±
5% of one of the 5 following values:
.0625 VDC, .25 VDC, 1 VDC, 4 VDC, 16 VDC.
<
range
>
may be specified in
millivolts (mv).
•
CALibration:CONFigure:RESistance
or
CALibration:CONFigure:VOLTage
commands
must be sent before
CALibration:VALue:RESistance
or
CALibration:VALue:VOLTage
commands.
•
The only
CALibration
command accepted after
CALibration:CONFigure:VOLTage
is
CALibration:VALue:VOLTage
.
•
The
CALibration:STORe
command may be used to store the calibration constants
in non-volatile memory.
•
Perform
ZERO
before
FSCale
or Error +3050 will be generated.
•
The
FSCale
output voltage of the calibration source will be greater than 90% of
the nominal value for each range, except the 16 V range where the output is 10 V.
•
This command will cause Error +3000, “ illegal while initiated” if trigger is initiated.
•
Related Commands:
CALibration:STORe ADC
,
CALibration:VALue:VOLTage
Command
Sequence
CAL:CONF:VOLT .0625,ZERO
Connect zero voltage reference to
Calibration Bus.
*OPC?
Must wait for CAL:CONF:VOLT to complete.
Now measure voltage with external DMM.
CAL:VAL:VOLT <
measured value
>
Inform HP E1413 of measured value.
CAL:CONF:VOLT .0625,FSCale
Connect full-scale voltage reference to
Calibration Bus.
*OPC?
Must wait for CAL:CONF:VOLT to complete.
Now measure voltage with external DMM.
CAL:VAL:VOLT <
measured value
>
Inform HP E1413 of measured value.
Repeat these activities for each of 4 other ranges (.25, 1, 4, 16).
.
.
.
CAL:STOR ADC
Store calibration constants in non-volatile
memory (use only at end of complete cal
sequence).
CALibration
178 HP E1313/E1413 Command Reference