Step 5. Setting the Sample Timer
The
SAMPle:TIMer <
scan_list
>,<
interval
>
command individually programs
channel-to-channel sample pacing. The power-on default allows maximum
sample rate.
•
The
<
scan_list
>
parameter can specify
LIST1
,
LIST2
,
LIST3
, or
LIST4
or
LISTL
. When
<
scan_list
>
is
LISTL
, the sample interval will
apply to all scan lists defined for
LISTL
while
ROUTe:SCAN
selects
LISTL
.
•
The
<
interval
>
parameter can specify
10.0e-6
(
MIN
) to
32.7680e-3
(
MAX
) seconds with a resolution of 0.5e-6 seconds.
Note
While the A/D filter is on (
[SENSe:]FILTer[:LPASs][:STATe] ON
) the
minimum sample time is 145
µ
sec. Use SCP filtering (filtering at channels)
to allow scanning at high speed.
To set the sample timer for scan list 1 to .005 seconds:
SAMP:TIM LIST1,5ms
5 ms for each channel in
scan list 1.