DIAGnostic:CUSTom:LINear
DIAGnostic:CUSTom:LINear <
table_range
>,<
table_block
>,(@<
ch_list
>)
downloads a custom linear Engineering Unit Conversion table (in
<
table_block
>
) to
the HP E1313/E1413. Contact your Hewlett-Packard System Engineer for more
information on Custom Engineering Unit Conversion for your application.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
table_range
>
numeric (float32)
.015625 | .03125 | .0625 | .125 | .25 | .5 |
1 | 2 | 4 | 8 | 16 | 32 | 64
volts
<
table_block
>
definite length block
data
See Comments below.
none
<
ch_list
>
channel list (string)
100 - 163
none
Comments
•
<
table_block
>
is a block of 8 bytes that define 4, 16-bit values. SCPI requires that
<
table_block
>
include the definite length block data header. C-SCPI adds the
header for you.
•
<
table_range
>
specifies the range of voltage that the table covers
(from –
<
table_range
>
to +
<
table_range
>
). The value you specify must
be within 5% of one of the nominal values from the table above.
•
<
ch_list
>
specifies which channels may use this custom EU table.
•
Related Commands:
[SENSe:]FUNCtion:CUSTom
•
*RST Condition: All custom EU tables erased.
Usage
Program puts table constants into array table_block.
DIAG:CUST:LIN
table_block
,(@116:123)
Send table to HP E1413 for channels 16 - 23.
SENS:FUNC:CUST:LIN 1,1,(@116:123)
Link custom EU with channels 16 - 23.
INITiate then TRIGger module.
DIAGnostic
190 HP E1313/E1413 Command Reference
Chapter 5