Trigger Timer and
Sample Timer Accuracy
100ppm (.01%) from -10
°
C to +70
°
C
External Trigger Input
TTL compatible input. Negative true edge triggered except first trigger will
occur if external trigger input is held low when module is INITiated. Minimum
pulse width 100nS. Since each trigger starts a complete scan of 2 or more
channel readings, maximum trigger rate depends on module configuration.
Maximum input voltage
(Normal mode plus common mode)
Operating: <
±
16 V peak Damage level: >
±
42 V peak
Maximum common mode
voltage
Operating: <
±
16 V peak Damage level: >
±
42 V peak
Common mode rejection
0 to 60Hz -105dB
Input impedance
greater than 100 Mohm differential
On-board Current Source
122
µ
A
±
0.02%, with
±
17 Volts Compliance
Maximum tare cal offset
SCP Gain = 1 (Maximum tare offset depends on A/D range and SCP gain)
A/D range
±
V F.Scale
16
4
1
0.25
0.0625
Max Offset
3.2213
.82101
.23061
.07581
.03792
Measurement accuracy
DC Volts
(90 days) 23
°
C
±
1
°
C (with *CAL? done after 1 hr warm up and CAL:ZERO?
within 5 min.). For HP E1313, multiply Noise Spec. by 1.4.
NOTE: When autorange in ON:
for
readings
<
3.8V,
add
±
0.02% to linearity specifications
for
readings
≥
3.8V, add
±
0.05% to linearity specifications
A/D range
±
V F.Scale
Linearity
% of reading
Offset Error
Noise
3 sigma
Noise*
3 sigma
.0625
.25
1
4
16
0.01%
0.01%
0.01%
0.01%
0.01%
5.3
µ
V
10.3
µ
V
31
µ
V
122
µ
V
488
µ
V
18
µ
V
45
µ
V
110
µ
V
450
µ
V
1.8 mV
8
µ
V
24
µ
V
90
µ
V
366
µ
V
1.5 mV
* [SENSe:]FILTer[:LPASs][:STATe] ON (min sample time - 145
µ
sec)
Temperature Coefficients: Gain - 10ppm/
°
C. Offset - (0 - 40
°
C) .14
µ
V/
°
C, (40 - 55
°
C) .8
µ
V+.38
µ
V/
°
C