Figure 5-1 shows the overall logical model of the Trigger System.
ARM[:IMMediate]
ARM[:IMMediate]
arms the trigger system when the module is set to the
ARM:SOURce BUS
or
ARM:SOURce HOLD
mode.
Comments
•
ARM[:IMMediate]
will cause Error -221, “ Settings Conflict” unless
ARM:SOURce
is
BUS
or
HOLD
.
•
Related Commands:
ARM:SOURce
•
*RST Condition:
ARM:SOURce IMMediate
Usage
ARM:IMM
After INIT, system is ready for trigger event.
ARM
Same as above (:IMM is optional).
ARM:SOURce
ARM:SOURce <
source
>
configures the AR
M
system to respond to the specified
source.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
source
>
discrete (string)
BUS | EXT | HOLD | IMM | SCP |
TTLTrg<
n
> (<
n
> = 1 through 7
none
Comments
•
The following table explains the possible choices.
Parameter Value
Source of Arm
BUS
ARM[:IMMediate]
EXTernal
Trig signal on terminal module.
HOLD
ARM[:IMMediate]
IMMediate
The arm signal is always true (continuous arming).
SCP
SCP Trigger Bus (future HP or SCP Breadboard).
TTLTrg<
n
>
The VXIbus TTLTRG lines (<
n
> = 0 through 7).
•
See the note about the
ARM
subsystem on page 160.
ARM
Chapter 5
HP E1313/E1413 Command Reference 161
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com