The Arm and Trigger Sources can be changed while the module is scanning (but
not in the continuous mode). This provides a way to stop triggering when you
want to change the
TRIGger:TIMer[:PERiod] <
trig_interval
>
. To do this, execute:
ARM:SOURce IMMediate
Must be IMMediate when SOURce not TIMer.
TRIGger:SOURce BUS
or
TRIGger:SOURce HOLD
To stop scanning.
TRIGger:TIMer <
new_interval
>
Change the timer interval.
TRIGger:SOURce TIMer
Return to timed scans.
•
When
TRIGger:SOURce
is
TIMer
,
TRIGger:TIMer:MODE
is set to
SYNC
, and
INITiate:CONTinuous
is
ON
, an
ARM
event is required only to trigger the first scan.
After that, the timer continues to run and the module goes to the wait-for-trigger state,
ready for the next timer trigger. An
INITiate:CONTinuous OFF
command will return
the module to the trigger idle state after the current scan is completed. See
TRIGger:SOURce
and
TRIGger:TIMer:MODE
for more details.
•
While
ARM:SOURce
is
IMMediate
, you need only
INITiate
the trigger system to
start a measurement scan.
•
Related Commands:
ARM[:IMMediate]
,
ARM:SOURce?
,
INITiate[:IMMediate]
,
INITiate:CONTinuous ON
,
TRIGger:SOURce
•
*RST Condition:
ARM:SOURce IMMediate
Usage
ARM:SOURce BUS
Arm with ARM command.
ARM:SOURce TTLTRG3
Arm with VXIbus TTLTRG3 line.
ARM:SOURce?
ARM:SOURce?
returns the current arm source configuration. See the
ARM:SOURce
command for more response data information.
•
Returned Value: Discrete, one of
BUS
,
HOLD
,
IMM
,
SCP
, or
TTLT0
through
TTLT7
. The C-SCPI type is string.
Usage
ARM:SOUR?
An enter statement return arm source
configuration.
ARM
162 HP E1313/E1413 Command Reference
Chapter 5