Usage
SENS:STRAIN:EXC? (@107)
Query excitation voltage for channel 7.
Enter statement here.
Returns the excitation voltage set by
STR:EXC.
[SENSe:]STRain:GFACtor
[SENSe:]STRain:GFACtor <
gage_factor
>,(@<
ch_list
>)
specifies the gage factor
to be used to convert strain bridge readings for the channels specified by
<
ch_list
>
.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
gage_factor
>
numeric (float32)
1 - 5
none
<
ch_list
>
channel list (string)
100 - 163
none
Comments
•
<
ch_list
>
must specify the channel used to sense the bridge voltage, not the
channel position on a Bridge Completion SCP.
•
Related Commands:
[SENSe:]STRain:GFACtor?
,
[SENSe:]FUNCtion:STRain:
…
commands
•
*RST Condition: Gage factor is 2.
Usage
STRAIN:GFAC 3,(@100:107)
Set gage factor for channels 0 through 7.
[SENSe:]STRain:GFACtor?
[SENSe:]STRain:GFACtor? (@<
channel
>)
returns the gage factor currently set
for the sense channel specified by
<
channel
>
.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
channel
>
channel list (string)
100 - 163
none
Comments
•
<
channel
>
must specify a single channel only.
•
Returned Value: Numeric value of gage factor. The C-SCPI type returned is flt32.
•
Related Commands:
[SENSe:]STRain:GFACtor
Usage
STRAIN:GFAC? (@107)
Query gage factor for channel 7.
Enter statement here.
Returns the gage factor set by STR:GFAC.
[SENSe:]
Chapter 5
HP E1313/E1413 Command Reference 249