Test 3. Sweep
Time Accuracy
Step 6. Sweep Time Accuracy, Sweep Times
ms
[
S W E E P T I M E
)
Marker A Time
Min
Measured
Max
20 ms
18 ms
22 ms
50 ms
45 ms
55 ms
100 ms
90 ms
110 ms
500 ms
450 ms
550 ms
900 ms
1.10
Step 12. Sweep Time Accuracy, Sweep Times
s
Step 19. Sweep Time Accuracy, Sweep Times
ms (Alternate Procedure)
[
S W E E P T I M E
)
Gen Readout
20 ms
50 ms
100 ms
500 ms
10
50
100
150
Min
18.0 ms
45.0 ms
90.0
450 ms
900 ms
9.00 s
45.0 s
90.0
20.0
Measured
2-72 Performance Tests