7. 3
Bandwidth Filter Adjustments
10 Hz Amplitude Adjustments
37.
38.
39.
40.
41.
42.
43.
Connect CAL OUTPUT to RF INPUT. Key in
PRESET
),
9,
10 Hz.
Adjust the instrument front panel FREQ ZERO control for
maximum signal amplitude on the CRT display.
Key in
1
and DISPLAY LINE
U
s
i
n
g
t
h
e
DATA knob, place the display line at the signal trace.
Key in
10 Hz.
Adjust the instrument front panel FREQ ZERO control for
maximum signal amplitude on the CRT display.
Adjust
10 Hz AMPTD and
10 Hz AMPTD equal
amounts to set the signal level the same as the reference level set
in step 39. See Figure 3-38 for location of 10 Hz AMPTD adjusts.
Repeat steps 37 through 42 until no further adjustment is
required.
3-76 Adjustments
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...