4. Final Display Adjustments (SN
and Above)
4. Final Display
Adjustments (SN
and Above)
Reference
Description
Equipment
Procedure
Note
Keyboard
X, Y, Z Axis Amplifiers
This procedure is used to optimize the appearance of the CRT display
during routine maintenance or after CRT replacement or minor
repairs. First, the display is adjusted for best focus over the full CRT,
then the graticule pattern is adjusted for optimum rectangular display.
Digital Photometer . . . . . . . . . . . . . . . . . . . . . . . .
Option 02
Photometer Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
56503
Photometer interconnect cable . . . . . . . . . . . . . . . Tektronix 012-0414-02
Photometer light
. . . . . . . . . . . . . . . . . . . .
016-0305-00
Adjustment Procedure 2, High Voltage Adjustment (SN
and
Above) should be performed prior to performing the following
adjustment procedure.
1. Connect the equipment as shown in Figure 3-31.
ANALYZER
Figure 3-31. Final Display Adjustments Setup
2. Set the photometer probe to NORMAL. Press
on the
photometer to turn it on and allow 30 minutes warm-up. Zero the
photometer according to the manufacturer’s instructions.
3. With the spectrum analyzer’s LINE switch set to STANDBY,
set the potentiometers listed in the
3-6 as indicated. See
Figure 3-32 for the location of the adjustments.
Note
In this procedure, do not adjust the following potentiometers and
variable capacitors on the
X, Y, Z Amplifier Assembly:
or
These components are
adjusted in the factory and in Adjustment Procedure 3, Preliminary
Display Adjustments (SN
and Above).
Adjustments
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...