10. IF Gain Uncertainty Test
2
Gain Steps
13. Press
7.
14. Set
[
REFERENCE
to -1.9
15. Press MARKER (OFF). Set
to 100 Hz.
16. Set the frequency synthesizer for an output power level of -3.9
Set the amplitude increment for 2
steps.
17. Press MARKER
[
PEAK
18. Set the analyzer
LEVEL
)
and the frequency synthesizer
amplitude according to
2-13. At each setting, note the MKR
A amplitude and record it in the table.
2-13. IF Gain Uncertainty, 2 Steps
LEVEL
)
Frequency
Deviation
Synthesizer (MARKER A
- 1 . 9
- 3 . 9
- 5 . 9
- 7 . 9
- 9 . 9
Amplitude
‘Amplitude
- 3 . 9
0 (ref)
- 5 . 9
- 7 . 9
- 9 . 9
- 1 1 . 9
2 . 0 0
Figure 2-19. IF Gain Uncertainty Measurement (2
2-36 Performance Tests
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...