8. Frequency Response Test
100
32.
33.
34.
35.
36.
37.
38.
39.
40.
41.
Press
PRESET
)
on the spectrum analyzer. Activate SIGNAL
INPUT 1.
Key in the following spectrum analyzer settings:
START
. . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
. . . . . . . . . 100
Connect equipment as shown in Figure 2-15 with function
generator to SIGNAL INPUT 1.
Set the function generator controls as follows:
LINE . . . . . . . . . . . . . . . . . . .. . . . . . .. . . . . . . . . . . . . . . . . . . . ON
RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 K
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL (button in)
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 V
AMPLITUDE VERNIER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . midrange
SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
TRIGGER PHASE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FREE RUN
MODULATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . all out
MODULATION RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I
M O D U L A T I O N R A N G E H z V E R N I E R . . . . . . . f u l l y C C W
MODULATION SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
Percent Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CW
Adjust function generator FREQUENCY to place signal between
the last two graticule lines (right side) on the signal analyzer
display.
Adjust AMPLITUDE VERNIER on the function generator until
the peak of the signal is at the reference graticule line on the
spectrum analyzer display.
Press LOG
1
on the spectrum analyzer. Press
DISPLAY LINE
and set the Display Line to the level
recorded for 100
in step 25.
Adjust function generator AMPLITUDE VERNIER to place peak of
signal at the Display Line.
Adjust FREQUENCY on the function generator to position the
signal trace at the right edge of the spectrum analyzer display
(last graticule line).
Press MODULATION SWP on the function generator and allow the
function generator to make at least two complete sweeps. Press
TRACE A [MAX HOLD). Allow the function generator to make one
complete sweep. After completion of the sweep, press TRACE A
Performance Tests
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...