12. Option 857 Amplitude Fidelity
Frequency
1
Synthesizer Calibrated
Amplitude Amplitude
Step
0 (ref)
0
- 1 0
- 1 0
-20
-20
-30
-30
-40
-40
-50
-50
-60
-60
-70
-70
-80
-80
-90
20. Subtract the greatest negative fidelity error from the greatest
positive fidelity error for calibrated amplitude steps from -10
to -70
The results should be
21. Subtract the greatest negative fidelity error from the greatest
positive fidelity error for calibrated amplitude steps from -10
to -90
The results should be
5-2. Log Amplitude Fidelity (10
Option 857)
2
Fidelity Error
MARKER A Amplitude (Column 2 Column 1)
0 (ref)
0 (ref)
Cumulative Cumulative
Error
Error
0 to 80
0 to 90
Linear Fidelity
22. Key in analyzer settings as follows:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300
Hz
FREQUENCY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10
23. Set the frequency synthesizer for an output power level of + 10
24. Press SCALE LIN pushbutton. Press MARKER
[
PEAK
to center the signal on the display.
25.
Set
(
FREQUENCY SPAN
)
to 0 Hz and
to 1 Hz. Press (SHIFT),
(resolution bandwidth), MARKER
26. Decrease frequency synthesizer output amplitude by 10
steps,
. noting the MARKER A amplitude and recording it in column 2 of
5-3.
Option 857
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...