4. Impulse and
Resolution
Bandwidth
Accuracy
Related Adjustment
Impulse Bandwidth Adjustments
Specification
3 MHz bandwidth
1 MHz to 1
bandwidths
-0,
300 Hz to 10 Hz (6
bandwidths)
Description
A frequency synthesizer and pulse/function generator are used
to input pulses to the spectrum analyzer. The amplitude of the
pulses is measured, and the impulse bandwidths are calculated for
each impulse bandwidth from 3 MHz to 1
The 6
resolution
bandwidths are then measured using the spectrum analyzer
function. The CAL OUTPUT signal is used as a stable signal source to
measure the 6
resolution bandwidths.
SPECTRUM ANALYZER
SYNTHESIZER
LEVEL GENERATOR
.
PULSE/FUNCTION
GENERATOR
.
J
L
Figure 4-2. Impulse Bandwidth
Setup
Equipment
Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Pulse/Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Procedure
1. Set the frequency synthesizer for a 15 MHz, + 13
output.
Connect the output of the frequency synthesizer to the EXT
INPUT of the pulse/function generator.
2. Set the pulse/function generator controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TRIG
EXT INPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . positive-going
EXT INPUT LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . midrange
OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . pulse
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OV
HIL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
WIDTH (WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 ns
DISABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
4-4 Option 462
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...