4. Final Display Adjustments (SN
and Above)
Figure 3-32. Location of Final Display Adjustments on
3-6. Initial Adjustment Positions
Adjustment
Y GAIN
Y POSN
X GAIN
X POSN
INT GAIN
FOCUS COMP
T/B FOC
Al
T/B CTR
Al
R/L FOC
A
R/L CTR
ORTHO
INT LIM
ASTIG
Front-panel INTENSITY
Front-panel FOCUS
Front-panel ALIGN
Position
centered
centered
centered
centered
two-thirds clockwise
centered
centered
centered
centered
centered
centered
centered
fully counterclockwise
centered
fully counterclockwise
centered
centered
4. Set the spectrum analyzer’s LINE switch to ON, and wait at least
5 minutes to allow the CRT and high-voltage circuits to warm up.
5. Set the front panel INTENSITY control fully counterclockwise and
adjust
INT LIM until the display is just visable. See
Figure 3-32.
6. Set the front-panel INTENSITY control fully clockwise.
7. Adjust
X GAIN,
X POSN,
Y
GAIN, and
Y POSN for optimum centering of the
display annotation and graticule pattern.
3-62 Adjustments
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...