3. Preliminary Display Adjustments (SN
and Above)
12. Set the oscilloscope controls as follows:
Press
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
amplitude scale
offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..25.000 0 V
Channel 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
amplitude scale
offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..60.000 0 V
Press (TRIG):
source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
.
level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25.0000 V
Press
time scale . . . . . . . . . . . . . . . . . . . . . . .
50.0
delay . . . . . . . . . . .. . . . . . . . . . . . . .. . . . . . . . . . . . . . . 125.000 ns
Press (DISPLAY):
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on
Press
13. Set the spectrum analyzer’s front-panel INTENSITY control fully
counterclockwise, and then set the LINE switch to ON.
14. The X+ deflection and X- deflection waveforms should be
superimposed on the oscilloscope display, as shown in Figure 3-26.
If necessary, adjust
X POSN and
X GAIN
for a centered display of at least four vertical divisions. See
Figure 3-25 for the location of the adjustments.
running
- 1 2 5 . 0 0 0
1
1 0 . 0
2 5 . 0 0 V
10.00 :
4
1 0 . 0
6 0 . 0 0 v
10.00
25.00 V
Figure 3-26. X + and X- Waveforms
15. Set the oscilloscope controls as follows:
Press
f l . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
display . . . . . . . . . . . .. .. .. .. .. . .. .. .. . . . . . . . . . . . . . . . . . . .. .. on
math . . . . . . . . . . . . . . . .. . . . . . . . . . . channel 1 channel 4
sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...25.0
16. Three waveforms should be displayed on the oscilloscope, as
shown in Figure 3-27. The lower composite waveform represents
Adjustments 3-55
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...