1. Center Frequency Readout Accuracy Test
Equipment
Synthesized Sweeper . . . . . . . . . HP
Frequency Standard . 10 MHz standard,
within 1 part in
e.g. HP
Adapter, Type N (m) to SMA (f) . . . HP1250-1250
61 cm (24 in.) Cable Assembly, SMA Male Connectors HP 5061-1086
Procedure 1.
2 .
3 .
4 .
5 .
6 .
7 .
Connect CAL OUTPUT to SIGNAL INPUT 2.
Press
on the analyzer.
Adjust FREQ ZERO for a maximum amplitude trace.
Press
Set the synthesized sweeper for a 100.000 MHz signal at a level of
approximately 0
Connect equipment as
shown in
Figure
Set analyzer
FREQUENCY
)
and
[
FREQUENCY SPAN
)
and
synthesized sweeper frequency according to Table 2-2. At each
setting, press
[
PEAK
to center the signal. Adjust
LEVEL
)
as necessary to place signal peak at a convenient
level.
Record the CENTER readout frequency in the table for each
setting. The limits for this frequency are given in the table. See
Figure 2-2.
Figure 2-2. Center Frequency Readout Error Measurement
2-4 Performance Tests
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...