4. Impulse and Resolution Bandwidth Accuracy Test (p/o
2-19, Performance Test Record)
Steps 39 through 46. 6
Resolution Bandwidth
Accuracy
Res
BW
3 MHz (i)
1 MHz (i)
300
(i)
(i)
30
(i)
10
(i)
3
(i)
1
(i)
300 Hz (i)
100 Hz (i)
30 Hz (i)
10 Hz (i)
5 MHz
2 MHz
500
200
50
20
5
2
500 Hz
200 Hz
100 Hz
100 Hz
300 Hz
100 Hz
30 Hz
10 Hz
450 Hz
150 Hz
45 Hz
15 Hz
4-20 Option 462
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...