13. 20 MHz Reference Adjustments
Spectrum Analyzer
(
CENTER FREQUENCY
_)
to 20.34 MHz and
SCALE to 10 dB/division.
10. Adjust
20.34 MHz NULL
for minimum 20.34 MHz
signal at
as indicated by HP
Spectrum Analyzer
display. With signal
the plates of the NULL adjustment
capacitor should be meshed approximately halfway. If fully
meshed or fully unmeshed, a circuit malfunction is indicated.
11. Disconnect frequency synthesizer from
and reconnect cable
2 (red) to
Connect power meter to rear-panel INT REF
OUT connector.
12. Power meter indication should be no more than 5
less than
that noted in step 3
Time Base output).
13. Disconnect
from ground. Connect power meter to
14. Adjust
COMB DRIVE
for power meter indication of
+ 10.0
15. Connect power meter to
through cable 3 (orange). Power
meter indication should be at least -15
Reconnect cable 3
(orange) to
16. Connect power meter to
through cable 4 (yellow). Power
meter indication should be at least -10
Reconnect cable 4
(yellow) to
17. On the oscilloscope, key in
to perform a soft
reset.
18. Connect
the
channel 1 probe to the oscilloscope’s rear panel
PROBE COMPENSATION AC CALIBRATOR OUTPUT connector.
Press
[
AUTO SCALE
).
Adjust the channel 1 probe for an optimum
square wave display on the oscilloscope.
19. Connect oscilloscope with the HP
probe to
and
the ground to the analyzer’s chassis ground.
20. Set the oscilloscope controls as follows:
Press
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . .
div
offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..OV
coupling
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..dc
probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
Press
EDGE TRIGGER . . . . . . . . . . . . . . . . . . . . . . . . trig’d auto
source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 800 mv edge
Press
time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 ns
delay . . . . . . .
reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..CNTR
Press
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . .......... AVG
Adjustments
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...