21. Slope Compensation Adjustments
Note
At this sweep time, some trace discontinuities are common.
14. Adjust
TILT for best flatness (clockwise rotation increases
the power slope), and trigger two sweeps on the synthesized
sweeper. See Figure 3-81 for the location of
Compare
the resultant trace with the specification. Continue adjusting
until best flatness is achieved.
Note
Best flatness is achieved when the maximum number of frequency
points are on or near the -14
reference.
F R E O U E N C Y C O N T R O L
I
T I L T
Figure 3-81. Location of
Adjustment
15. Press TRACE A,
[
PEAK SEARCH
],
and
[
MARKER DELTA
).
Using
the data knob, place the marker on the lowest power peak. The
marker’s absolute value should be less than 2
16. See Figure 3-82 for examples of typical displays of frequency
response correctly and incorrectly adjusted.
FREQUENCY RE
L I M I T S
PROPERLY
TED
Figure 3-82. Slope Compensation Adjustment Waveforms
Adjustments 3-135
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...