3. Preliminary Display Adjustments (SN
and Below)
Procedure
X and Y Deflection
1. Connect a 10:1 (10
divider probe to
the
oscilloscope’s channel
Amplifier Pulse
1 input and a 10: 1 divider probe to the channel 4 input.
Response Adjustments
2. On
the
oscilloscope, press
to perform a soft reset.
On the oscilloscope, press
more
preset probe ,
select
channel 1, and use the front-panel knob to select a
probe.
4. Select channel 4, and use the front-panel knob to select a
probe.
5. Press
6. Connect the channel 1 probe to the oscilloscope’s rear panel
PROBE COMPENSATION AC CALIBRATOR OUTPUT connector.
Press
[
AUTO
-
Adjust the channel 1 probe for an optimum
square wave display on the oscilloscope.
7. Connect the channel 4 probe to the oscilloscope’s rear panel
PROBE COMPENSATION AC CALIBRATOR OUTPUT connector.
Press
SCALE
_).
Adjust the channel 4 probe for an optimum
square wave display on the oscilloscope.
Note
Each probe is now compensated for the oscilloscope input to which it
is connected. Do not interchange probes without recompensating.
8. Connect
the
channel 1
divider probe to
the
channel 4 probe to
as shown in Figure 3-16. Connect
the probe ground leads to chassis ground. See Figure 3-17 and
Figure 3-18 for the location of the assemblies and test points.
PULSE/FUNCTION
GENERATOR
Figure 3-16. Preliminary Display Adjustments Setup
9. Remove the cover over
Digital Storage Section and remove
Intensity Control Assembly. Insert the Display Adjustment
PC board (HP part number 85662-60088) into the
slot. See
Figure 3-17 for the location of the
assembly.
3-46 Adjustments
Содержание 8568B
Страница 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Страница 19: ......
Страница 37: ......
Страница 42: ......
Страница 48: ......
Страница 53: ......
Страница 54: ...9 RF Gain Uncertainty Test h I 1 I I I Figure 2 16 RF Gain Uncertainty Measurement 2 32 Performance Tests ...
Страница 61: ......
Страница 90: ......
Страница 100: ...Test 9 RF Gain Uncertainty Test Step 6 2nd LO Shift Min Measured Max 1 0 dB 1 0 dB 2 78 Performance Tests ...
Страница 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Страница 109: ...Test 16 Residual FM Test Test 16 Residual FM Test Step 14 Residual FM IMinIMeasuredIrax Performance Tests 2 87 ...
Страница 162: ......
Страница 224: ......
Страница 268: ......
Страница 270: ......
Страница 272: ......
Страница 278: ......
Страница 307: ...w43 c c5 c4 c3 c2 Figure 6 l RF Section Top View A26 6 4 Major Assembly and Component locations ...
Страница 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...