6.3 Performing Statistical Calculations on Measurement Values
85
6
When statistical calculation is set to ON and an EXT. I/O trigger signal is applied, operation will be
performed as follows depending on the trigger source setting:
• With external [EXT] triggering: One measurement is performed and subject to statistical calculation.
• With internal [INT] triggering: The next measurement value after the trigger signal is subjected to statistical
calculation.
Operation is the same in the following cases:
Using Statistical Calculations
(Key Operations)
(Remote Control)
• When the
F4
[MANU]
key on the [MENU] - [TRG] selection
screen is pressed.
•
∗
When a
TRG
remote control command is
received.
• When the
F2
[PRINT]
key on the Measurement screen is
pressed (with internal triggering and Auto-Memory disabled.
Appears only when the interface is set for the printer.).
• When an EXT. I/O PRINT signal is applied
on the Measurement screen (with internal
triggering and Auto-Memory disabled).
• When acquiring measurement values with the Auto-Memory
function (p. 81).
1
Open the Basic Settings screen.
2
Open the Data Memory Settings screen.
3
Enable or disable the function.
4
Return to the Measurement screen.
The Basic Settings screen appears.
The Data Settings screen appears.
[DATA SETTINGS]
1
Selection
2
Selection
1
2
Disables statistical calculation (default).
Enables statistical calculation.
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previ-
ous screen.
The confirmation screen appears.
When statistical calculation is enabled,
F3 [STAT]
appears on the Measurement screen.
Confirm calculation results (p. 86)
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......