![Hioki RM3542-50 Скачать руководство пользователя страница 86](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361086.webp)
6.1 Storing Data at Specified Timing (Data Memory Function)
80
Measurement values are stored in the instrument's internal memory at the following timings (up to
30,000 points).
• All measurements performed by external (EXT) triggering
• When a trigger is applied during internally (INT) triggered measurement
The following three storage methods are available:
• Stores when an EXT. I/O TRIG signal is received (p. 95)
• Stores when a
∗
TRG
command is received (p. 138)
• Stores when the F4 [MANU] key on the [MENU] - [TRG] setting screen is pressed.
• This function can only be enabled by remote control. The data memory function should be enabled by
remote command beforehand. This setting is not available on the front panel.
• Stored memory data cannot be viewed on the instrument's screen. Use remote control commands to export
stored data.
Data Memory Function Operating Procedure
Moreover, stored data is automatically erased at the following timings:
6.1 Storing Data at Specified Timing
(Data Memory Function)
1
Enable data memory mode.
Send this remote command to enable the data memory function:
:MEMory:MODE MEMory
2
Store measurement values.
Execute external trigger measurement, or apply a trigger during internally triggered measurement.
3
Export the stored data.
Send this remote command to export the measurement values stored in the instrument:
:MEMory:DATA?
4
Clear measurement data from instrument memory.
Send this remote command to erase the data from instrument memory:
:MEMory:CLEar
•
When the memory function setting (including auto-mem-
ory) is changed (p. 156)
•
When the range is changed (p. 32)
•
When changing comparator settings (p. 36)
•
When printing the statistical calculations (p. 93)
•
When the measuring object is changed (p. 28)
•
Upon system reset (p. 75)
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......