4.3 Compensating Measured Values (Scaling Function)
46
When measuring the resistance of current sensing resistors, there may be a discrepancy between
the resistance value at the time of mounting on the board being used and the resistance value
obtained when measuring the component alone (due to the effects of the probing location and other
factors). The scaling function corrects the measured resistance value obtained from the component
alone to yield the resistance value during actual use. Scaling is performed by mean of the following
equations:
R
S
=
A
×
R
+
B
R
:
Measured value before compensation
R
S:
Resistance value after compensation
A
:
Compensation coefficient setting range: 0.50000 to 2.00000 (default:: 1.00000)
B
:
Offset resistance setting range: ±0.0000 m
Ω
to ±99.9999 M
Ω
(default: 0.0000 m
Ω
)
• Scaling calculation will be performed on a measured value after zero-adjustment. Consequently, measure
value may not be equal to zero after zero-adjustment.
• Changing the comparator settings automatically sets the scaling function to OFF.
4.3 Compensating Measured Values
(Scaling Function)
1
Open the Basic Settings screen.
2
Open the Measurement Settings Screen.
3
Enable or disable the scaling function.
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Disables the function
(default)
(go to step 6)
Enables the function
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......