9.8 Data exporting methods
163
9
Basic Data Exporting Methods
Flexible data exporting is available depending on the application.
9.8 Data exporting methods
Free-Run Data Exporting
Default setting
:INITiate:CONTinuous ON
(continuous measurement enabled)
:TRIGger:SOURce IMMediate
(internal trigger)
Exporting
:FETCh?
Exports the most recent measurement.
Data Export by Host Trigger
Default setting
:INITiate:CONTinuous OFF
(continuous measurement disabled)
:TRIGger:SOURce IMMediate
(internal triggering)
Exporting
:READ?
A trigger occurs, and a measurement is taken and the result is transferred.
Exporting by pressing the F4 [MANU] Key*
*1
or applying the TRIG signal
Default setting
:INITiate:CONTinuous OFF
(continuous measurement disabled)
:TRIGger:SOURce EXT
(external trigger)
Exporting
:READ?
When a TRIG signal is input by pressing the
F4
[MANU] key
*1
or TRIG signal, measures
and transfers the result.
*1. Press the
F4
[MANU] key displayed when [TRG: EXT] has been set on the Basic Settings Screen.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......