8.2 Timing Chart
99
8
Each signal level indicates a corresponding voltage level.
•
This shows the operations when the trigger source setting is [EXT] and the EOM output setting is Pulse.
•
A self calibration measurement (approximately 130 ms) is automatically performed every 10 minutes. (between mea-
surements). During that time, a TRIG signal input is accepted but the corresponding measurement is delayed until self
calibration is finished.
•
Do not apply a TRIG signal while measuring using external triggering (the signal is ignored).
•
When changing settings, such as the measurement range, allow for processing time (approximately 150 ms) before
applying a TRIG signal.
•
Input signals are ignored when the following are displayed: Basic, Detailed, and Comparator Settings screens; Statistical
Calculation Results screen (except for the PRINT signal); and error messages (except for the Setting Monitor errors).
•
The INDEX signal OFF timing is the same as that of the EOM signal OFF timing.
•
The output of the comparator judgments and error signals are determined before the EOM signal turns ON.
•
With internal triggering, [INT], the EOM signal is fixed at High (OFF). Also, comparator judgments and error signals are
not de-asserted (OFF) when measurement starts.
EOM operation when the EOM Output setting is Hold
At the end of the measurement, the EOM signal is asserted ON.
Until the next TRIG signal is input and measurement starts again, the EOM signal continues to be asserted.
See:
"Setting the End-of-Measurement Signal Output (EOM Signal Setting)" (p. 105)
When the next measurement is started by accepting the next TRIG signal, the EOM signal is de-asserted (OFF).
8.2 Timing Chart
Contact check
Measurement
currents
Contact Improver
Measurement
processing
Checking
End of
Measurement
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......