3.7 Judging Measurement Values (Comparator Function)
37
3
The comparator function is enabled by default.
When the function is disabled, comparator settings are ignored.
Example: Set a reference value of 10.5
Ω
with +4.5% and -4.5% judgment tolerances.
To abort the setting process, press
. Settings are abandoned and the display returns to the previous
screen.
Enabling and Disabling the Comparator Function
1
Open the Basic Settings screen.
2
Enable or disable the comparator function.
3
Return to the Measurement screen.
The Basic Settings screen appears.
1
Selection
2
Disables the function.
Enables the function.
Press the up/down
cursor keys to change
the setting.
(When the function is disabled)
Comparator judgments are indicated only when the function is enabled.
Decide According to Reference Value and Tolerance (REF% Mode)
1
Open the relative tolerance setting screen.
Relative value =
Reference
value
Measurement
value
- 1
×
100 [%]
Setting range:
-9.999% to +9.999% (When 10% or less)
-99.99% to +99.99% (When more than 10%)
Reference value
Positive tolerance (upper judgment thresh-
old)
Negative tolerance (lower judgment thresh-
old)
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......