9.7 Message Reference
138
Write and Read the Service Request Enable Register (SRER)
Syntax
Command
∗
SRE
<0 to 255 (NR1)>
Query
Response
∗
SRE?
<0 to 255 (NR1)>
Description
Command
The SRER mask is set to the numerical value 0 to 255. Although NRf
numerical values are accepted, values to the right of the decimal are rounded
to the nearest integer.
Bit 6 and unused bits 2, 3 and 7 are ignored.
The data is initialized to zero at power-on.
Query
The contents of the SRER, as set by the
∗
SRE
command, are returned as an
NR1 value (0 to 255). Bit 6 and unused bits 2, 3 and 7 are always returned as
zero.
Example
∗
SRE 33
Set SRER bits 0 and 5 to 1.
∗
SRE?
33
SRER bits 0 and 5 have been set to 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
unused
0
ESB
MAV
unused
unused
ESE1
ESE0
Read the Status Byte and MSS Bit
Syntax
Query
Response
∗
STB?
<0 to 255 (NR1)>
Description
The contents of the STB are returned as an NR1 value (0 to 255).
The response message has no header.
Example
∗
STB?
16
STB bit 4 has been set to 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
unused
MSS
ESB
MAV
unused
unused
ESE1
ESE0
Request a Sample
Syntax
Command
∗
TRG
Description
Performs one measurement when external triggering is enabled.
When Statistical Calculation is ON, calculation data is exported.
When the memory function is ON, the measurement value is stored.
Example
:TRIGger:SOURce EXTernal;
∗
TRG
Note
When an error occurs with the Settings Monitor function enabled, triggering is disabled
(p. 59).
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......