9.5 Communication Methods
120
Compound Command Header Omission
When several commands having a common header are combined to form a compound com-
mand (e.g.,
:CALCulate:LIMit:REFerence
and
:CALCulate:LIMit:PERCent
),
if they are written together in sequence, the common portion (here,
:CALCulate:LIMit:
)
can be omitted after its initial occurrence.
This common portion is called the "current path" and, until it is cleared, the interpretation of
subsequent commands assumes that they share the same common portion (in which the
"current path" has been omitted).
This usage of the current path is shown in the following example:
Full expression
:CALCulate:LIMit:REFerence
1.0E+3;:CALCulate:LIMit:PERCent 1.0,-1.5
Compacted expression
:CALCulate:LIMit:REFerence 1.0E+3;PERCent 1.0,-1.5
The current path is cleared when the power is turned on, when reset by key input, by a colon
":" at the start of a command, and when a message terminator is detected.
Standard command messages can be executed regardless of the current path.
They have no effect upon the current path.
A colon ":" is not required at the start of the header of a Simple or Compound command. How-
ever, to avoid confusion with abbreviated forms and operating mistakes, we recommend that
a colon is always placed at the start of a header.
Output Queue
Response messages are stored in the output queue until read by the controller and then
cleared. The output queue is also cleared in the following cases:
• Power on
• Device clear
• Query error
The output queue capacity of the instrument is 64 bytes. If response messages overflow the
buffer, a query error is generated and the output queue is cleared.
Also, with GP-IB, if a new message is received while data remains in the output queue, the
output queue is cleared and a query error is generated.
Input Buffer
The input buffer capacity of the instrument is 256 bytes.
If 256 bytes are accumulated in the buffer and it becomes full, the GP-IB interface bus enters
the waiting state until space is cleared in the buffer.
The RS-232C interface cannot accept data beyond 256 bytes.
This portion becomes the current path, and can be omitted from the messages
immediately following.
Output Queue and Input Buffer
Ensure that the commands are maintained below 256 bytes.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......