4.4 Setting the Measurement Integration Time Option
49
4
•
The instruments accuracy specifications are applicable only with the default integration times.
Investigate your measurement requirements carefully before changing the integration time.
•
When the effects of power line noise can be ignored, the integration time can be set longer than the default to reduce
scattering of measurement values. On the other hand, if the integration time is too short, scattering increases.
For high-or low-resistance and low-power resistance measurements that are easily affected by power line noise, we
suggest setting according to the power line period (PLC units).
3
Select the integration setting units.
4
Select the integration time.
5
Return to the Measurement screen.
Selection
1
Sets in units of time.
Sets in units of power line cycles.
2
The setting is specific to the selected range (p. 43).
Selection
1
Setting range:
•
When setting in ms units:
0.1 ms to 100.0 ms
•
When setting power-line-cycle units:
1 to 6 PLC (60 Hz), 1 to 5 PLC (50 Hz)
2
3
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previ-
ous screen.
The confirmation screen appears.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......