9.7 Message Reference
141
9
(3) Sampling
Execute and Clear Zero-Adjustment
Clear Zero-Adjustment
Syntax
Command
:ADJust:CLEar
Description
Command
Clears any zero-adjustment.
Execute Zero-Adjustment
Syntax
Query
Response
:ADJust?
<0/ 1>
0 = Indicates that zero-adjustment has succeeded.
1 = Indicates that the offset resistance value exceeded 10
Ω
during zero-adjustment.
Set and Query Measurement Speed
Syntax
Command
:SPEEd
<FAST/ MEDium/ SLOW
>
Query
Response
:SPEEd?
<FAST/ MEDIUM/ SLOW>
Example
:SPEE MED
:SPEE?
MEDIUM