Chapter 10 Specifications
184
(7) Faulty Measurement Detection
Out-of-Range Detection Function
Operation
Indicates under- or over-range values in the following conditions:
•
Measurement value is outside of the measurement range
•
The relative calculation value is outside of the display range
•
Measurement value is outside of the A/D converter input range
•
The zero-adjusted value is outside of the display range
Ex.: Zero-adjustment in the 1
Ω
range with 0.5
Ω
zero-adjustment in effect, measur-
ing 0.1
Ω
gives a calculation result of -0.4
Ω
, which is outside of the display range.
Contact Check Function
Operation
Checks the connections between H
POT
and H
CUR
, and between L
POT
and L
CUR
termi-
nals (for each range)
Default setting
ON, 200
Ω
Setting
ON/ OFF
Threshold setting
50
Ω
/ 100
Ω
/ 150
Ω
/ 200
Ω
/ 300
Ω
/ 400
Ω
/ 500
Ω
(reference value)
Current Monitor Function
Operation
Detects faults that can obstruct normal measurement current. This function cannot be
disabled.
Voltage Level Monitor Function
Operation
Detects an error when the detection voltage is unstable (for each range)
Default setting
ON, NORMAL (LOOSE is the default only for the 100 M
Ω
range)
Setting
ON/ OFF/ ALLOFF
Threshold setting
LOOSE
NORMAL
SEVERE
Comparator Judgment and Display Indications
Table 3. Measurement States
and Display Indications
Current monitor
PASS
FAIL
Contact check
voltage level monitor
PASS
Measurement value is
displayed
Normal judgment
Out-of-range display
Hi/Lo judgment
FAIL
Contact Error
No judgment
Contact Error
No judgment
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......