![Hioki RM3542-50 Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361065.webp)
4.10 Comparing the Measurement Settings of Two Instruments (Settings Monitor Function)
59
4
This function automatically compares the settings of two
instruments to determine whether they are the same.
Only comparator settings and measurement speed are
compared.
When the settings differ, an alarm notification is displayed
and subsequent.
TRIG
signal input is prevented from starting measure-
ment.
When the settings of two instruments match, TRIG input is
accepted and measurement starts.
However, if the range defined by the upper and lower thresh-
olds of the second stage is broader than that of the first stage,
measurement still starts despite the different threshold settings.
4.10 Comparing the Measurement Settings of
Two Instruments (Settings Monitor Func-
tion)
Automatic
Comparison
1st Stage
2nd Stage
Transport
Direction
A
B
C
D
E
1
Connect the SET MONITOR connectors of the two instruments using a Hioki 9637 RS-
232C cable.
2
Open the Basic Settings screen.
3
Open the System screen.
4
Enable or disable the function. Enable the function on both instruments.
The SET MONITOR connectors are identical
to RS-232C connectors.
Be careful to avoid connecting the wrong
connectors.
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
The setting is not monitored (default) (go
to step 7).
The setting is monitored.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......