9.7 Message Reference
153
9
Set and Query the Continuous Measurement Setting
Syntax
Command
:INITiate:CONTinuous
<1/ 0/ ON/ OFF>
Query
Response
:INITiate:CONTinuous?
<ON/ OFF>
<ON> = Continuous Measurement Enabled
<OFF> = Continuous Measurement Disabled
Description
•
Continuous Measurement Enabled:
After measurement, enters the Trigger Wait State. When the trigger source setting is
IMMediate, the next trigger occurs immediately (the Free-Run State).
•
Continuous Measurement Disabled:
After measurement, enters the Idle State instead of the Trigger Wait State.
•
Triggering is ignored in the Idle State. Executing :INITiate[:IMMediate] enables the
Trigger Wait State.
•
Continuous measurement is enabled upon exit from the Remote State.
Example
:INIT:CONT OFF
:INIT:CONT?
ON
Trigger Wait Setting
Syntax
Command
:INITiate[:IMMediate]
Description
Switches triggering from the Idle State to the Trigger Wait State.
Example
Disable continuous measurement, and read one value for each trigger event.
Sending
:TRIG:SOUR IMM
........Trigger occurs immediately after the system enters the Trigger
Wait State
:INIT:CONT OFF
........Disables continuous measurement
:INIT
...............................Setting the system in the Trigger Wait State execute the triger
immediately since :TRIG:SOUR IMM has been set
:FETC?
............................. Fetch the measurement value
Receiving
2.3
.................. Measurement value is 2.16414 k
Ω
Error
•
When continuous measurement is enabled (:INITIATE:CONTINUOUS ON), an execution
error occurs.
Note
•
When the trigger source is IMMediate, triggering occurs immediately before entering the
Idle State.
•
When the trigger source is EXTernal, the Trigger Wait State is enabled to wait for an
external trigger, and when a trigger occurs, one measurement is taken before entering
the Idle State.
Trigger Source Setting and Queries
Syntax
Command
:TRIGger:SOURce
<IMMediate/ EXTernal>
Query
Response
:TRIGger:SOURce?
<IMMEDIATE/ EXTERNAL>
<IMMEDIATE> = Internal triggering
<EXTERNAL> = External triggering
Description
Press the F4 [MANU] key on the [MENU]-[TRG] selection screen (Basic settings screen) to
trigger by the TRIG signal or
∗
TRG
command.
Example
:TRIG:SOUR IMM
:TRIG:SOUR?
IMMEDIATE
Note
When the external [EXT] is set to ON, the auto-memory function will be disabled (OFF).
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......