6.3 Performing Statistical Calculations on Measurement Values
86
Statistical calculation results are displayed on the screen.
Printing is available using commercially available printers having a serial interface. Statistical calcu-
lation results are automatically erased after printing.
Before printing, select the [PRINT] interface setting.
See:
"7.2 Setting the Instrument" (p. 91)
The number of valid samples can be confirmed on the Statistical Calculation Results screen.
• When the number of valid samples (Val) is zero, calculation results are not displayed.
• When only one valid data sample exists, no standard deviation or process capability indices are displayed.
Confirming, Printing, and Erasing Statistical Calculation Results
1
Display the Calculation Results screen.
2
To print
To print, select the printer as the interface setting on the System screen (p. 91).
To erase
Calculation Results screen is displayed
(only if statistical calculation is
enabled).
Num
Total data count
Val
Number of valid measurement values (error-free data) (Valid)
Ave
Mean
Sn
Population standard deviation
Max
Maximum
Sn1
Standard deviation of sample
Min
Minimum
Cp
Process capability index (variation)
Cpk
Process capability index (bias)
Output to the printer.
"Example Printouts (PRINT MODE:
NORMAL)" (p. 94)
Statistical calculation results and stored data are erased when printing finishes.
Erases the last measurement and calcula-
tion result (executes only once).
Erases all measurement values and statisti-
cal calculation results.
After selecting, a confirmation screen ap-
pears.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......