298
Application Measurement
>
>
4
4
1
1
22
33
44
55
66
4
Under
[P1]
in
[Pulse]
, configure the settings as follows:
1
1
Input type
Count
22
Count mode
Addition
33
Slope
↑ (depends on the specifications of watt-meter)
44
Threshold
1 V (depends on the specifications of watt-meter)
55
Filter
ON *
66
Scaling
Decimal, 1 kWh = 5000 (5 k), Unit: kWh
*To prevent false counting caused by chatter.
You can use the scaling function to convert the pulse count to energy (kWh).
10000 kW
0 kW
5
Press the
START
key to start measurement.
Data will be recorded at a 30 min. interval for 30 days, and waveform data will be saved on the SD
Memory Card.
Recording will stop 30 days after it starts.
If you wish to end recording sooner, press the
STOP
key.