96
Activating a Trigger at a Set Interval
OR and AND conditions
Trigger
conditions
: Trigger point;
: Other trigger condition
OR
Interval time
Interval time
Measurement start
AND
Measurement start
Interval time
Interval time
The trigger will not
activate.
There is one enabled trigger during the interval.
The trigger will not activate if it does not match other trigger conditions. (If no other trigger
conditions have been set, this setting will result in the same operation as OR logic.)