108
Saving Data
3
Under
[Format]
in the
[Waveform data]
area, select the auto save type.
OFF
Does not perform auto-save operation.
Binary
Saves data in the instrument’s dedicated format (binary format)
Text
Saves data in text format (.CSV).
(When
[Format]
is set to
[Text]
)
Under
[
Downsampling]
, select whether to perform downsampling.
When set to
[ON]
, the size of the saved data file will be reduced.
OFF
,
ON
(When
[Downsampling]
is set to
[ON]
)
Under
[Decimation factor]
, select the decimation factor.
1/2
to
1/100,000
Example: If set to
[1/5]
, 1 data point out of every 5 data points will be left.
Under
[Deleting]
, select the type of processing to perform if the media on which data is
being saved runs out of space (as necessary).
OFF
Ends auto-save operation.
ON
Deletes the oldest waveform file and continues auto-save operation. If the system is
unable to delete a file, it will stop saving data. Numerical calculation results will not be
deleted.
Under
[Folder splitting]
, select the timing at which to segment data (as necessary).
The folders in which data is saved can be segmented based on the set period of time.
Disable
Saves in a single folder such files containing data obtained during the amount of time
set for recording. The folder will be segmented under the following conditions:
• When
[File splitting]
is set to
[Disable]
:
Segments the folder by day.
•
When the number of files in the folder has reached 1000:
Switches the save destination over to folders with a sequential number appended to
(for example, 19-12-23_0001).
1 d
Segments folders by day.
1 week
Segments folders by week. A week is defined as lasting from Monday to Sunday.
1 month
Segments folders by month.
Example: If set to
[1 week]
When measurement started on Sunday, December 29, 2019, the system will regard Monday,
December 23, 2019 as the beginning of the week.
The
19-12-23
folder will automatically be created.
Example: If set to
[1 month]
When measurement started on December 29, 2019, the system will regard December 1, 2019 as
the beginning of the month.
The
19-12-01
folder will automatically be created.
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...