295
Application Measurement
>
>
[Module n]
(n = 1, 2, . . .)
7
7
1
1
22
7
Configure the settings as follows in the
[Input]
area for the channel being measured:
1
1
Input type
Voltage
22
Range
1-5 V
Since a 250
Ω
resistor has been connected to the input terminals, 4 mA will be recorded as 1 V, and 20 mA
will be recorded as 5 V.
The 1-5 V range is a range in which the 10 V range’s display range lower and upper limits are set to 1 V and 5 V,
respectively.
If you wish to change the upper and lower limit values, use 10 V range.
5 V
1 V
8
Press the
START
key to start measurement.
Data will be recorded at a 5 s interval for one hour.
In addition, the average value numerical calculation will be performed every minute and the results
saved to the SD Memory Card.
Recording will stop 1 hr. after it starts.
If you wish to end recording sooner, press the
STOP
key.
11
Knowledge and Information