11
Setting Measurement Conditions
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
>
>
2
2
3
3
4
4
5
5
6
6
1
1
1
Enter a title comment in the
[Title comment]
field (optional).
2
Under
[Recording mode]
, select type of recording.
Normal
Data is recorded in synchronization with the internal clock.
The setting cannot be changed from
[Normal]
. Hioki plans to add new functionality in the future.
3
Under
[Recording interval]
, select the data capture interval.
Example: Choosing
[10 ms]
will cause data to be captured at a 10 ms interval (100 times per
second).
1 ms
*
1
,
2 ms
*
1
,
5 ms
*
1
,
10 ms
,
20 ms
,
50 ms
,
100 ms
,
200 ms
,
500 ms
,
1 s
,
2 s
,
5 s
,
10 s
,
20 s
,
30 s
,
1 min
,
2 min
,
5 min
,
10 min
,
20 min
,
30 min
,
1 h
*1: Setting available only when using a module with data refresh intervals including 1 ms. (when there is at
least one U8553, U8554, LR8533, or LR8534 module for which measurement is enabled).
4
Under
[Repetitive recording]
, choose whether to repeat recording operation.
OFF
Stops recording after one iteration.
ON
Repeats recording operation.
Measurement will stop when you press the
STOP
key.
1
Settings and Operation
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...