Configuring External Input/Output (I/O) Terminals
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Trigger output timing
The timing at which signals appear in trigger output varies with the
[Timing]
setting.
See “2.2 Enabling the Trigger Function” (p. 84).
Start
Measurement start
Start trigger activated
Measurement stop
= Acquisition complete
To next
measurement
Minimum
pulse width
Trigger output waveform
10 ms or greater
Level held during
measurement
•
Trigger output will switch to active when the start trigger activates.
•
The pulse will be output for at least 10 ms and held while measurement continues.
•
Trigger output will switch to non-active when measurement stops.
Stop
Measurement start
Stop trigger activated
= Acquisition complete
To next
measurement
Minimum
pulse width
Trigger output waveform
10 ms or greater
•
Trigger output will switch to active when the stop trigger activates.
•
The pulse will be output for at least 10 ms, and then trigger output will switch to non-active.
Start & stop
Measurement start
Start trigger activated
Stop trigger activated
= Acquisition complete
To next
measurement
Minimum
pulse width
Trigger output waveform
10 ms or greater
Level held during
measurement
•
Trigger output will switch to active when the start trigger activates.
•
The pulse will be output for at least 10 ms and held while measurement continues.
•
Trigger output will switch to non-active when the stop trigger activates.
8
External Control (EXT
. I/O)
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...